Remote Control Commands
R&S
®
SMB100A
306Operating Manual 1407.0806.32 ─ 13
Usage: SCPI confirmed
Manual operation: See "Current Freq - Frequency Sweep" on page 152
[:SOURce<hw>]:FREQuency:MODE <Mode>
The command sets the instrument operating mode and therefore also the commands
used to set the output frequency.
Parameters:
<Mode> CW | FIXed | SWEep | LIST
CW|FIXed
The instrument operates in fixed-frequency mode; CW and
FIXed are synonyms. The output frequency is set
with :SOURce:FREQuency:CW|FIXed.
SWEep
The instrument operates in SWEep mode. The frequency is set
using the commands SOURce:FREQuency:STARt; STOP;
CENTer; SPAN; MANual.
LIST
The instrument processes a list of frequency and level settings for
the selected path.
The List mode settings are made in the SOURce:LIST Subsys-
tem subsystem. The setting SOURce:FREQuency:MODE LIST
also sets the command SOURce:POWer:MODE automatically to
LIST.
Correlation: FREQ:MODE LIST sets POW:MODE LIST
*RST:
CW
Example:
FREQ:MODE SWE
sets the SWEep mode. The settings under
SOURce:FREQuency:STARt; STOP; CENTer; SPAN; MANual
become effective.
Manual operation: See "State - Frequency Sweep" on page 149
[:SOURce<hw>]:FREQuency:MULTiplier <Multiplier>
The command sets the value for the multiplication factor of a subsequent downstream
instrument.
Parameters:
<Multiplier> float
Range: 0.001 to (100 GHz / current RF frequency)
*RST: 1
Example:
FREQ:MULT 1
sets the multiplication factor to 1.
Manual operation: See "Multiplier - RF Signal" on page 118
SOURce Subsystem