EasyManua.ls Logo

Rohde & Schwarz SMB100A - Page 178

Rohde & Schwarz SMB100A
481 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Function
R&S
®
SMB100A
162Operating Manual 1407.0806.32 ─ 13
Execute Single - List Mode
Triggers the list manually. This button is available only if mode "Single" is selected.
Remote command:
[:SOURce<hw>]:LIST:TRIGger:EXECute on page 332
Reset - List Mode
Resets the list to the starting point.
Remote command:
[:SOURce<hw>]:LIST:RESet on page 331
Dwell Time - List Mode
Enters the dwell time. The dwell time determines the duration of a list step in list operating
modes "Auto", "Single" and "Extern Single". In these modes a complete list is processed
either once or continuously.
In list operating modes "Step" and "Extern Step", the set dwell time does not affect signal
generation. In this case, the duration of a list step is determined by the time between two
(internal or external) trigger events.
The "Dwell Time" set by the user is used as the step time of the list mode. The effective
net dwell time is shorter, reduced by the setting time. This setting time may be greater
than the time specified in the data sheet.
Remote command:
[:SOURce<hw>]:LIST:DWELl on page 327
Current Index - List Mode
Sets the list index in "Step" mode.
Remote command:
[:SOURce<hw>]:LIST:INDex on page 328
Learn List Mode Data... - List Mode
Starts the determination of the hardware setting for the selected list. The data determined
in this way is stored along with the list.
It may be necessary to deliberately activate list learning in the event of greatly altered
environmental conditions that require new hardware settings.
RF Block

Table of Contents

Other manuals for Rohde & Schwarz SMB100A

Related product manuals