Instrument Function
R&S
®
SMB100A
193Operating Manual 1407.0806.32 ─ 13
"Sawtooth"
One sweep runs from start to stop frequency. Each subsequent sweep
starts at the start frequency, i.e. the shape of the sweep sequence
resembles a sawtooth.
"Triangle"
One sweep runs from start to stop frequency and back, i.e. the shape
of the sweep resembles a triangle. Each subsequent sweep starts at
the start frequency.
Remote command:
[:SOURce]:LFOutput:SWEep[:FREQuency]:SHAPe on page 316
Step Lin/Log - LF Sweep
Sets the step width for the individual sweep steps. This entry is effective for all sweep
modes.
"Step Lin" or "Step Log" is displayed depending on whether "Spacing Lin" or "Log" is
selected.
"Step Lin"
With the linear sweep, the step width is a fixed frequency value which
is added to the current frequency. The linear step width is entered in
Hz.
Remote command:
[:SOURce]:LFOutput:SWEep[:FREQuency]:STEP[:LINear] on page 317
"Step Log"
With the logarithmic sweep, the step width is a constant fraction of the
current frequency. This fraction is added to the current frequency. The
logarithmic step width is entered in %.
Remote command:
[:SOURce]:LFOutput:SWEep[:FREQuency]:STEP:LOGarithmic on page 318
Dwell Time - LF Sweep
Sets the dwell time. The dwell time determines the duration of the individual sweep steps.