Remote Control Commands
R&S
®
SMB100A
391Operating Manual 1407.0806.32 ─ 13
:STATus:QUEStionable:NTRansition <Ntransition>
Sets the bits of the NTRansition part of the STATus:QUEStionable register. If a bit is set,
a transition from 1 to 0 in the condition part causes an entry to be made in the EVENt
part of the register.
Parameters:
<Ntransition> string
Example:
STAT:OPER:NTR 0
a transition from 1 to 0 in the condition part of the Status:Ques-
tionable register does not cause an entry to be made in the EVENt
part
:STATus:QUEStionable:PTRansition <Ptransition>
Sets the bits of the NTRansition part of the STATus:QUEStionable register. If a bit is set,
a transition from 1 to 0 in the condition part causes an entry to be made in the EVENt
part of the register.
Parameters:
<Ptransition> string
Example:
:STAT:OPER:PTR 32767
all transitions from 0 to 1 in the condition part of the Status:Ques-
tionable register cause an entry to be made in the EVENt part
STATus:QUEue[:NEXT]?
Queries the oldest entry in the error queue and then deletes it. Positive error numbers
denote device-specific errors, and negative error numbers denote error messages
defined by SCPI. If the error queue is empty, 0 ("No error") is returned.
The command is identical to :SYSTem:ERRor[:NEXT]? on page 394.
Return values:
<Next> string
Example:
:STATus:QUEue?
queries the oldest entry in the error queue.
Response: 0, 'no error'
no errors have occurred since the error queue was last read out
Usage: Query only