Remote Control Commands
R&S
®
SMB100A
390Operating Manual 1407.0806.32 ─ 13
:STATus:PRESet <Preset>
Resets the status registers. All PTRansition parts are set to FFFFh (32767), i.e. all tran-
sitions from 0 to 1 are detected. All NTRansition parts are set to 0, i.e. a transition from
1 to 0 in a CONDition bit is not detected. The ENABle parts of STATus:OPERation and
STATus:QUEStionable are set to 0, i.e. all events in these registers are not passed on.
Parameters:
<Preset> string
Example:
STAT:PRES
resets the status registers.
:
STATus:QUEStionable:CONDition <Condition>
Queries the content of the CONDition part of the STATus:QUEStionable register. This
part contains information on the action currently being performed in the instrument. The
content is not deleted after being read out since it indicates the current hardware status.
Parameters:
<Condition> string
Example:
:STATus:QUEStionable:CONDition?
queries the Status:Questionable:Condition register.
:STATus:QUEStionable:ENABle <Enable>
Sets the bits of the ENABle part of the STATus:QUEStionable register. This setting
determines which events of the Status-Event part are enabled for the sum bit in the status
byte. These events can be used for a service request.
Parameters:
<Enable> string
Example:
STAT:OPER:ENAB 1
problems when performing an adjustment cause an entry to be
made in the sum bit.
:STATus:QUEStionable[:EVENt] <Event>
Queries the content of the EVENt part of the STATus:QUEStionable register. This part
contains information on the actions performed in the instrument since the last readout.
The content of the EVENt part is deleted after being read out.
Parameters:
<Event> string
Example:
STAT:QUES:EVEN?
queries the Status:Questionable:Event register.
STATus Subsystem