52Operating Manual 1407.0806.32 ─ 13
f) Finally, select "State" and press the rotary knob to switch on the RF frequency
sweep.
g) Press the DIAGRAM key to display the complete block diagram.
The "RF" is not yet active, which means that no RF signal is output.
3. Activate RF signal.
a) Turn the rotary knob to select the "RF" block.
b) Press the RF ON/OFF key to activate the "RF" signal output.
To indicate the active state, the RF block is displayed in blue. An RF signal with the
default frequency and level settings is output, i.e. 1 GHz and -30 dBm.
The sweep is not yet active, it must be triggered in the sweep dialog.
4. Trigger RF frequency sweep
a) Press the WINBAR key to switch to the "RF Frequency Sweep" dialog.
Turn the rotary knob to select the "Execute Single Sweep" button.
Press the rotary knob to trigger (start) the frequency sweep.
A linear single sweep signal is now present at the RF output, starting at 100 MHz.
The sweep is processed in 1 MHz steps with dwell time of 10 ms per step up to the
stop frequency of 500 MHz.
The sweep starts at 100 MHz, stops at 500 MHz in 1 MHz steps is output with a dwell
time of 10 ms per step.