Program Mode 6-19
Disabled:
No correction pressure applied.
Factor:
The correction method applied is:
Corrected flow = measured flow + ((test pressure - measured pressure) * factor).
Absolute:
The correction method is based on ratio of absolute pressures:
Corrected flow = measured flow * (test pressure + atmosphere) / (measured
pressure + atmosphere)
Gauge:
The correction method is based on the ratio gauge pressures:
Corrected flow = measured flow * (test pressure / measured pressure)
1.) Measured flow is the reading at the end of the measure step. 2.) Measured
pressure is the reading at the end of the measure step. 3.) Test pressure is the
target pressure entered on the specification tab. 4.) Atmosphere = 14.7psi.