<6. Parameter Lists>
212
IM 01E21A02-03EN
Relative
Index
Index
Parameter
Name
Default Value
Write
Mode
Description
AI1 AI2 AI3
AI4
(only
AXG)
23 4023 4123 4223 4323 ACK_OPTION 0xFFFF AUTO
Sets operation to acknowledge
(acknowledgmentforthe
alarm)ofvariousalarms.
By setting a bit to the alarm,
the device operates for the
alarm as if it is acknowledged
without the acknowledgment.
Bit assignment complies with
the Communication standard
specications.
24 4024 4124 4224 4324 ALARM_HYS 0.5 AUTO
Threshold(hysteresis)against
clearing each alarm of HI_HI,
HI, LO and LO_LO. An active
alarm is cleared if the setpoint
goes into the normal value side
by the amount of hysteresis
from the limit value.
25 4025 4125 4225 4325 HI_HI_PRI 0 AUTO
Speciesthepriorityforthe
HI_HI alarm. It can be used
not just to set the priority, but
todisablealarmnotication.
Meanwhile, to activate an
alarm on communication, it is
necessary to specify the priority
for alarm activation in advance.
To activate an alarm, set 3 or
more. By sending with priority
information being added in a
communication frame when the
alarm is activated, for example,
the alarm is used to apply a
ltertoignorevaluesbelowthe
speciedpriorityonthehost
side.
26 4026 4126 4226 4326 HI_HI_LIM FLT_MAX AUTO
Speciesalimitvalueforthe
HI_HI alarm.
27 4027 4127 4227 4327 HI_PRI 0 AUTO
SpeciesthepriorityoftheHI
alarm. Meanwhile, to activate
an alarm on communication,
it is necessary to specify the
priority for alarm activation in
advance. To activate an alarm,
set 3 or more. By sending with
priority information being added
in a communication frame
when the alarm is activated, for
example, the alarm is used to
applyaltertoignorevalues
belowthespeciedpriorityon
the host side.
28 4028 4128 4228 4328 HI_LIM FLT_MAX AUTO
SpeciesalimitvaluefortheHI
alarm.