10-10 Maintenance
General Maintenance 5DX Series 3
Procedure:
1. Bring up the X-rays to operating nominals from the User Interface Menu.
2. Select System Access and press Enter.
3. Select Startup and press Enter.
4. Select Automatic and press Enter.
5. Select HV Power Status and press Enter.
Figure 10-2: HV Power Status Window
6. If your system has a printer connected, press Shift and the Print Screen Key on
the keyboard. This will print the information displayed on the User Interface
Monitor. If there is no printer connected to the 5DX System, then perform the
following steps to capture the information for printing on another computer.
a. Press Alt + Print Screen keys simultaneously.
b. Click on Start. Click on Programs. Click on Accessories. Click on
Paint.
c. Click on Edit. Click on Paste.
127(
The path can be to a floppy disk drive or networked drive that will
give you access to a computer with print capabilities.