10-44 Maintenance
Software Diagnostics 5DX Series 3
Run Resolution Test
Maintenance Frequency: Monthly
Estimated Time: Five minutes
Required Items: None
Information: The procedure described below is a manual test for the 400 FOV
only. Agilent recommends that you run the automated resolution tests procedure
by selecting Utilities > Toolbox from the User Interface menu. The Agilent 5DX
Toolbox contains fully automated Confirmation and Diagnostic tests that provide a
simple pass or fail indication. The purpose of these tests is to confirm that your
system hardware is operational or help diagnose failures quickly when it is not.
Procedure:
1. Select Imaging > Image Resolution from the Service menu and press
Enter.
2. The resolution test runs and measures the resolution of the system. It displays
the resolution value as an MTF percent. For information on pass/fail criteria,
run the automated resolution test procedure outlined in the 5DX Service
Quick Reference Guide.