10-26 Maintenance
Main Cabinet Maintenance 5DX Series 3
Check the System Air Pressure/Low Pressure Regulator
Maintenance Frequency: Monthly
Estimated Time: Ten minutes
Required Items:
None
Prerequisite:
None
Special Notes:
None
Information:
None
Procedure:
1. Locate the Pneumatic Gauge, Regulator and Filter Assembly at the rear of the
Main Cabinet.
2. Check the Pneumatic Gauge and verify that the pressure reading is set to 80psi.
3. If the pressure reading is not 80psi, then adjust the regulator until the Pneu-
matic Gauge reads 80psi.
a. Pull up on the regulator control valve.
b. Turn the regulator control valve clockwise or counter-clockwise until the
correct reading is obtained.
c. Push down on the regulator control valve after the adjustment is complete.
4. Locate the Low Pressure regulator assembly on left rear of the Main Cabinet.
a. Check that the gauge is reading approximately 40psi.
b. If the gauge reading is less than 40psi or higher than 50psi, adjust the low
pressure regulator for the 40psi.
c. If the pressure cannot be adjusted, contact the Agilent Response Center.

Table of Contents

Other manuals for Agilent Technologies 5DX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5DX and is the answer not in the manual?

Agilent Technologies 5DX Specifications

General IconGeneral
BrandAgilent Technologies
Model5DX
CategoryMedical Equipment
LanguageEnglish

Related product manuals