6-14 Pneumatic Subsystem
Pneumatic Subsystem Components 5DX Series 3
Adjusting the Air Cylinder
1. At the Command Prompt, type cshtr and press Enter. This command closes
the Inner Barrier.
2. At the Command Prompt, type auto and press Enter. This will start the User
Interface.
3. Select Utilities and press Enter.
4. Select Service Menu and press Enter.
5. Select Panel Handling and press Enter.
6. Select Hardware Control and press Enter.
7. Set the Hardware Status and Control to Input mode by pressing I (the letter).
8. Open one of the Outer Barriers by typing either 10 for the Left Outer Barrier
or 12 for the Right Outer Barrier. Open the access panels and perform the
required adjustment. (See Figure 6-8)