EasyManuals Logo

Agilent Technologies 5DX Service Guide

Agilent Technologies 5DX
276 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #263 background imageLoading...
Page #263 background image
Agilent 5DX Service Guide 10-39
5DX Series 3 Main Cabinet Maintenance
Clean and Inspect the Main Cabinet
Maintenance Frequency: Monthly
Estimated Time: Ten minutes
Required Items:
Vacuum Cleaner
Clean, lint-free and non-abrasive cleaning wipes
#2 Phillips screwdriver
Prerequisite:
Shutdown the 5DX System. Follow the Long Term Shutdown For the
Agilent 5DX System procedure in Chapter 4, Power Subsystem to ensure
you shut it down correctly.
Perform Lock-Out, Tag-Out procedure. See page 4-4 of Chapter 4, Power
Subsystem for more information.
Special Notes:
None
Information:
None
Procedure:
1. Open up all of the Main Cabinet access panels.
2. Vacuum the interior of the Main Cabinet.
3. Use wipes to clean up excess grease on the XYZ Stage Assembly and on the
interior of the Main Cabinet.

Table of Contents

Other manuals for Agilent Technologies 5DX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5DX and is the answer not in the manual?

Agilent Technologies 5DX Specifications

General IconGeneral
BrandAgilent Technologies
Model5DX
CategoryMedical Equipment
LanguageEnglish

Related product manuals