10-32 Maintenance
Main Cabinet Maintenance 5DX Series 3
5. Using the thickness gauge, check the gap between the XYZ Stage conveyor
belt to the bottom edge of flag. Adjust the gap as required.
a. Loosen the two screws that hold the flag in place.
b. Adjust the gap using the thickness gauge.
c. Retighten the screws.
6. Turn on the conveyor belts by clicking the XYZ Stage Belts On button in the
Hardware Status and Control window.
7. Verify the belt motion does not hit any of the flags.
8. Reverse the belt direction by clicking the Belt Direction Reverse button.
9. Verify again that the belts do not hit any of the flags.
10. Set the Belt Direction to Forward, and turn Off the XYZ Stage Belts.
11. Select Close from the File menu to exit the Hardware Status and Control
program.