GPIB Operation ML24xxA Native Commands
ML2437A/38A OM/PM PN: 10585-00001 Rev. P 8-11
8-8 ML24xxA Native Commands
This section provides an alphabetical listing of the GPIB commands (mnemonics) used to
program the Model ML243xA Power Meter in ML24XXA (native) mode. The emulation mode
can be set through the front panel SYSTEM > more > more > Rear panel > GPIB > MODE
menu (see Chapter 4, Operation) or through the GPIB command EMUL.
All ML24XXA (native) GPIB commands that use parameters must have a space between the
command header and the first parameter, and all subsequent parameters must be separated
by a comma (,). Multiple commands may be sent on the same line, but must be separated by a
semicolon (;).
The format for ML24XXA (native) GPIB commands is:
<command header><space><parameter 1>,<parameter n>,...
The end of the command text must be terminated with a line feed character (0Ah, decimal 10)
or a GPIB End of Transmission State (EOI), or both.
*CLS (Clear GPIB status bytes)
Syntax: *CLS
Remarks: This command performs a status data structure clear command. The event
status register and the status register are cleared except for the MAV bit.
*CLS does not clear the output buffer.