GPIB Operation ML24xxA Native Commands
ML2437A/38A OM/PM PN: 10585-00001 Rev. P 8-43
GRPIX (Profile type)
GRPRD (Profile data collection period)
GRPTP (Graph Pretrigger Percentage)
Syntax: GRPIX <mode>
mode: NORM, MINMAX, MIN, MAX, AVG
Remarks: Changes the type of graph displayed:
NORM: Profiles the sensor readings vs. time from the triggered point.
MINMAX: Plots both the MIN and MAX values for each point on the
graph. If connect points (GRCP) is ON, a vertical bar is drawn between
the min and max points.
MIN: Same as NORM, but each point is the minimum value that point
has achieved.
MAX: Same as NORM, but each point is the maximum value that point
has achieved.
AVG: This position plotted on the chart for an x-axis time slot is the
average of all the readings during that x-axis time slot period, and is
only available in Power vs. Time mode.
Query: GRPIX?
Returned String:
GRPIX <mode> (Mode can be AVG in Power vs. Time mode.)
Syntax: GRPRD <val>[units]
val: 100 ns to 7 s
Remarks: Sets the time the system will collect data for and scale into the profile
graph after a trigger event.
Example: GRPRD 20US sets the data collection period to 20 microseconds.
Query: GRPRD?
Returned String: GRPRD <val>
Syntax: GRPTP <val>[units]
val: 0 to 100
units: PCT
Remarks: Sets the pre trigger percentage of the profile screen. The percentage of
the data collection period that shows pretrigger information if the
display trigger delay is 1/PRF.
Query: GRPTP?
Returned String:
GRPTP <val>