HP438A Emulation Commands GPIB Operation
8-124 PN: 10585-00001 Rev. P ML2437A/38A OM/PM
DE (Display enable)
DR (Set GPIB address)
EMUL (GPIB emulation mode)
FA (auto average)
Syntax: DE
Remarks: Return the display to normal operation after the display has been set in
DD mode.
Related
commands:
DD
Syntax: DR<val>
val: 1 to 30
Remarks: Changes the device address when operating in HP emulation mode. The
power meter default address is 13.
Syntax: EMUL <mode>
mode:
ML24XX (Anritsu ML243xA native mode)
HP436A (Hewlett-Packard)
HP437B (Hewlett-Packard)
HP438A (Hewlett-Packard)
ML4803 (Anritsu ML4803A Series)
Remarks: Set the GPIB emulation to emulate other types of power meters. This
command is available in any emulation mode, and resets the whole
GPIB interface when the emulation mode is changed.
When selecting GPIB emulation modes, the instrument configures itself
to the preset conditions of the instrument to be emulated. For example,
when selecting HP438A emulation, the front panel menus pass through
the presets for the HP437B (which presets sensor A to dBm) then selects
HP438A emulation (which presets sensor A to Watts).
Syntax: FA
Remarks: Automatic Filter on. Allows the system to automatically select the filter
used to reduce the jitter in the display.
Related
commands:
FM, FH