GPIB Operation ML24xxA Native Commands
ML2437A/38A OM/PM PN: 10585-00001 Rev. P 8-61
OFFFIX Offset fixed value
OFFTBL (Specify the table used to apply offsets to the sensor)
Syntax: OFFFIX <s>,<val>[units]
s: A or B
val: -200.000 to +200.000
units: dB
Remarks: The value added to the sensor if the offset type is set to FIXED.
Example: To set the fixed offset for sensor A to -47 dBm:
OFFFIX A,-47DB
Query: OFFFIX? <s>
Returned String: OFFFIX <s>,<val>
Syntax: OFFTBL <s>, <val>
s: A or B
val: 1 to 5
Remarks: If the Offset Type is set to TABLE, use this command to specify which of
the five offset tables to apply to the sensor.
The tables are a set of frequency-against-dB offsets. The offset value
used from the table depends on the setting of the frequency correction
source. If the source is FREQUENCY, the entered frequency is used to
calculate the offset from the table. If the frequency correction source is
V/GHz, the frequency value calculated from the supplied ramp input is
used to calculate the offset from the table.
If the frequency does not match any frequency in the table, interpolation
is used to calculate the correct offset.
Note: If the frequency is greater than the maximum frequency in the table, the
offset value from the maximum table frequency is used. If the frequency
is less than the minimum frequency in the table, the offset from the
minimum table frequency is used. The frequency comparisons start from
the beginning of the table; if the entry is 0 Hz, this is counted as the end
of the table.
Query: OFFTBL? <s>
Returned String: OFFTBL <s>,<val>