GPIB Operation ML4803A Emulation Commands
ML2437A/38A OM/PM PN: 10585-00001 Rev. P 8-87
AVE (Sensor averaging setting)
CAL (Set the user cal factor value)
CCA (Clear the calfactor value to zero)
CDJ (Perform a CAL 0 dBm)
COF (Clear the offset value to zero)
COS (Turn ON the 50 MHz, 0 dBm RF calibrator output)
CRF (Clear the reference value to zero)
CST (Turn OFF the 50 MHz, 0 dBm RF calibrator output)
Syntax: AVE<number>
number:
0 = Averaging OFF
9 = HOLD. Holds the present averaged reading.
1 = Average for 1 s (ML243xA Repeat average number of 25).
2 = Average for 2 s (ML243xA Repeat average number of 70).
3 = Average for 5 s (ML243xA Repeat average number of 128).
4 = Average for 10 s (ML243xA Repeat average number of 256).
Remarks: The ML4803A averages for a period of time. The ML243xA sets the
averaging to repeat averaging for a number of readings.
Syntax: CAL<value>
value: Cal factor value in dB
Syntax: CCA
Syntax: CDJ
Remarks: During the cal 0 dBm sequence, the CAL execution bit in the status byte
is set. When the CAL operation is completed, the CAL execution bit is
cleared.
Syntax: COF
Syntax: COS
Syntax: CRF
Syntax: CST