ML24xxA Native Commands GPIB Operation
8-26 PN: 10585-00001 Rev. P ML2437A/38A OM/PM
CFUSEL (Select cal factor table)
CFUTBL (Number of cal factor tables in the sensor)
CFUUSE (Number of cal factor table being used)
CFUVLD (Valid cal factor table check)
Syntax: CFUSEL <s>,<table number>
s: A or B
table number: table number or combination to use
0 = factory default table
1 to 10 = user table being used
11 to 20 = factory table + user table being used
Remarks: Selects the cal factor table or combination of tables to be used and
automatically updates the sensor.
Example: CFUSEL A,13
Selects the factory table plus user table 3 in sensor A.
Syntax: CFUTBL<s>
s: A or B
Remarks: Returns the number of cal factor tables available in the selected sensor.
Syntax: CFUUSE<s>
s: A or B
Remarks: Returns a number indicating the cal factor table, or combination of
tables, being used by the selected sensor. Possible returned values are:
0 = factory default table
1 to 10 = user table being used
11 to 20 = factory table + user table being used
Syntax: CFUVLD <s>,<table number>
s: A or B
table number: 1 to number of tables supported by the sensor type
Remarks: Returns a TRUE if the table number passed is a valid initialized table
for the selected sensor. Returns a FALSE if it is not.