ML24xxA Native Commands GPIB Operation
8-60 PN: 10585-00001 Rev. P ML2437A/38A OM/PM
OBVSP (BNC analog output stop voltage scale)
OBVST (BNC analog output start voltage scale)
OBZL (BNC RF blanking output level when zeroing)
OFFCLR (Clear an offset table)
Syntax: OBVSP <port>,<val>
port: 1 or 2
val: -5.00 to +5.00 Volts
Remarks: Sets up the stop value for the voltage output in analog output mode.
Attempting to set the start value to a voltage greater than the stop
value, or the stop value lower than the start value, will result in an
execution error.
Query: OBVSP? <port>
Returned String: OBVSP <port>,<val>
Syntax: OBVST <port>,<val>
port: 1 or 2
val: -5.00 to +5.00 Volts
Remarks: Sets up the start value for the voltage output in analog output mode.
Attempting to set the start value to a voltage greater than the stop
value, or the stop value lower than the start value, will result in an
execution error.
Query: OBVST? <port>
Returned String: OBVST <port>,<val>
Syntax: OBZL <level>
level: HIGH (TTL high), LOW (TTL low)
Remarks: Sets the TTL level of the BNC RF blanking output.
Query: OBZL?
Returned String: OBZL <level>
Syntax: OFFCLR <val>
val: 1 to 5
Remarks: Sets all the values in the table specified to 0 dB and 0.00 Hz.