EasyManua.ls Logo

Anritsu ML2437A - OF (Offset State); OI (Return Identification String); OS (Set Offset Value); PR (Preset the Unit)

Anritsu ML2437A
314 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GPIB Operation HP437B Emulation Commands
ML2437A/38A OM/PM PN: 10585-00001 Rev. P 8-115
OF (Offset state)
OI (Return identification string)
OS (Set offset value)
PR (Preset the unit)
RA (Auto Range)
Syntax: OF<state>
state:
0 (OFF)
1 (ON)
Remarks: For example: OF1 (Turn offsets ON).
Syntax: OI
Remarks: Returned format:
<company name>,<model>,<firmware version>
Syntax: OS<val>EN
val: -200 to +200 (dBm only)
Remarks: Specifies the offset applied to the displayed value. Values can be entered
in 0.01 dB increments.
Example: OS10.13EN
Set an offset of 10.13 to the displayed value.
Syntax: PR
Remarks: Presets the unit to the HP factory defaults. This command does not
effect the calibration factors stored in the sensor data tables.
Syntax: RA
Remarks: Sets the ML243xA to automatically select the correct range for the
measurement.

Table of Contents