ML24xxA Native Commands GPIB Operation
8-34 PN: 10585-00001 Rev. P ML2437A/38A OM/PM
DUTY (Duty cycle)
DUTYS (Duty cycle state)
Syntax: DUTY <s>,<duty_cycle>
s: A or B
duty cycle: 0.1 to 100%
Remarks: Applies a duty cycle to the selected sensor. An offset will be applied
based on the entered value.
Example: DUTY A,50
Specifies a duty cycle of 50% that will alter the displayed readings by
approximately +3.01 dB.
Related
commands:
DUTYS
Query: DUTY? <s>
Returned String: DUTY <s>,<duty cycle>
Syntax: DUTYS <s>,<state>
s: A or B
state: ON or OFF
Remarks: Turns on or off the duty cycle for the selected sensor.
Related
commands:
DUTY
Query: DUTYS? <s>
Returned String: DUTYS <s>,<state>