GPIB Operation HP437B Emulation Commands
ML2437A/38A OM/PM PN: 10585-00001 Rev. P 8-107
CL (Cal Adjust)
CS (Clear all status byte)
CT (Clear the cal factor table)
DA (Display All)
Syntax: CL<val><terminator>
val: 50.0 to 120.0
terminator:
%
PCT
EN
Remarks: Same as the ML24XXA (native) CFADJ command. Sets a calibration
factor to be used when performing a 0 dBm calibration.
Examples:
CL98.5EN
CL98.5%
CL98.5PCT.
Syntax: CS
Remarks: Same as the *CLS command. Resets all of the GPIB status registers and
clears the input queue.
Syntax: CT<table_number>
table number: 0 to 9
Remarks: Clears the specified cal factor table to a single 50 MHz entry at 100%.
Since the ML243xA stores the cal factor table information in the sensor,
this data must be saved to the sensor by using the ‘EX’ command or an
additional command ‘SV’, or the data could be lost. The saving of the cal
factor table data to the sensor can be done at the end of all updates to a
particular table.
Syntax: DA
Remarks: Turns on all the segments of the display to verify proper operation. The
display is returned to normal when another command is sent.