8-70 PN: 10585-00001 Rev. P ML2437A/38A OM/PM
SENSTL (Set Sensor Settle Percentage)
SENTYP (Return sensor information)
SRCMOD (Source Sweep Mode)
SRCSPFRQ (Source Sweep Stop Frequency)
Syntax: SENSTL <s>,<val>
s: A or B
val: 0.01 to 10%
Remarks: Sets how long the system waits for the signal to settle. The value
parameter is only used in DEFAULT measurement sensor mode. The
settling time allows some control over the tradeoff between speed, and
the extent to which a measurement has settled to its final value.
Query: SENSTL? <s>
Returned String: SENSTL <s>,<mode>
Syntax: SENTYP <s>
s: A or B
Remarks: This command returns information on the selected sensor in string
format: <sensor type>,<sensor serial>.
For example: MA2442A, 003277.
Syntax: SRCMOD <mode>
mode: FREQ, OWER
Remarks: Determines whether the voltage sweep applied to the V/GHz analog
input on the rear panel is interpreted as a frequency or power sweep.
The x axis of the graph on the display will be labelled appropriately.
Query: SRCMOD?
Returned String: SRCMOD <mode>
Syntax: SRCSPFRQ <freq_value>[units]
freq value: 10 kHz to 122 GHz
Remarks: Determines the stop frequency when in frequency sweep mode.
Query: SRCSPFRQ?