ML24xxA Native Commands GPIB Operation
8-38 PN: 10585-00001 Rev. P ML2437A/38A OM/PM
FBEEP (Fail Beep On/Off)
FHOLD (Fail indicator Hold)
Syntax: FBEEP <c>,<state>
c: 1 or 2
state: ON or OFF
Remarks: When ON, causes an audio beep every time the limits for the selected
channel fail. If FBEEP is ON, and FHOLD is ON, whenever the limits
specified for the channel have been exceeded, a beep sounds once every
second until FHOLD is turned OFF, or the CLEAR key (CLR) is pressed.
The FAIL indication is not affected by the CLEAR key, and can only be
cleared by turning FHOLD off. If a limit fail happens again, the alarm
sounds again.
Related
commands:
FHOLD
Query: FBEEP? <c>
Returned String: FBEEP <c>,<state>
Syntax: FHOLD <c>,<state>
c: 1 or 2
state: ON or OFF
Remarks: If the high or low limits fail, and this setting is turned on, the fail status
continues until the command is turned off. All BNC outputs, beeps and
displays continue to be in the 'fail' state until after the OFF is received.
Related
commands:
FBEEP
Query: FHOLD? <c>
Returned String: FHOLD <c>,<state>