GPIB Operation ML24xxA Native Commands
ML2437A/38A OM/PM PN: 10585-00001 Rev. P 8-15
*RCL (Recall a stored setup)
*RST (Reset Device)
*SAV (Save configuration)
*SRE (Setup service request enable register)
Syntax: *RCL <val>
val:
1 to 10
Remarks: The ML243xA can store up to 10 instrument configurations for
convenient recall. The configuration parameters stored are:
Sensor Setup, Channel Setup, and Trigger Setup.
This command sets the ML243xA to a configuration previously stored in
memory locations 1 through 10. Trying to recall a setup that has not
been saved will set the execution bit in the event register (EXE in ESR).
Related
commands:
*SAV
Syntax: *RST
Remarks: Resets the ML243xA to the default configuration (see Appendix A,
Section A-3). Offset tables are not cleared. The GPIB address and
emulation settings are not changed, and the input queue, output queue,
and status registers on the GPIB are not cleared. This command
produces the same result as the front panel key sequence System >
Setup > PRESET > RESET.
Syntax: *SAV <val>
val:
1 to 10
Remarks: Saves the configuration of the power meter into the memory location
specified. Sensor Setup, Channel Setup, and Trigger Setup are saved
along with all other instrument parameters.
Related
commands:
*RCL
Syntax: *SRE <val>
val:
8-bit mask
Remarks: Sets the Service request enable register bits.