EasyManua.ls Logo

Anritsu ML2437A - ODT (Output the Current Calibration Factor, Offset, and Reference Level)8-91; OFF (Set Sensor Offset Value); OI? (Request Identity)

Anritsu ML2437A
314 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GPIB Operation ML4803A Emulation Commands
ML2437A/38A OM/PM PN: 10585-00001 Rev. P 8-91
ODT (Output the current calibration factor, offset, and reference level)
OFF (Set sensor offset value)
OI? (Request identity)
Syntax: ODT
Remarks: These are output as three separate messages in the output buffer, as
shown below:
CAL factor: 18 ASCII characters + <CR><LF>
OFFSET value: 18 ASCII characters + <CR><LF>
REFERENCE level: 19 ASCII characters + <CR><LF>
Note: When the ODT, OMR, and OPW data output commands are received
simultaneously, only the command which is received last is valid.
Figure 8-4. ODT Data Output Format
Syntax: OFF<value>
value: Offset value to add to the sensor reading.
Syntax: OI?
Remarks: Response: <ML4803>
1 5 10 15
CAL
dB
<CR><LF>
OFFSET
dB
<CR><LF>
REFERENCE
5 character data
dBm <CR><LF>
±
±
.
.
.

Table of Contents