GPIB Operation ML24xxA Native Commands
ML2437A/38A OM/PM PN: 10585-00001 Rev. P 8-47
GTDLY (Set profile trigger sample delay)
GTGW (Set profile trigger gate width)
GTLVL (Set profile trigger level)
Syntax: GTDLY <val>[units]
val: 0.0 to 1.0 s
Remarks: Sets the time delay after the display trigger delay to when the system
starts to take readings and displaying them. This point is represented
by the left most cursor.
Note: Changing the left most cursor or trigger delay time updates either the
cursor or the delay time value.
Query: GTDLY?
Returned String: GTDLY <val>
Syntax: GTGW <val>[units]
val: 100ns to 7.0 s
Remarks: Sets the time the system uses to perform whatever calculations are set
up. The time interval is represented by the space between the left most
cursor and the right most cursor. Changing either cursor, or the Gate
width value, will update both the cursors and the gate width value. The
default gate width value is 20 ms.
Query: GTGW?
Returned String: GTGW <val>
Syntax: GTLVL <val>
val: -30 to +20 dBm
Remarks: When the system trigger in profile mode is set to either INTA or INTB
(internal sensor A or B) it will trigger on a power level given by the
sensor. This command sets the level.
Related
commands:
GTSRC, GTTYP
Query: GTLVL?
Returned String: GTLVL <val>