HP436A Emulation Commands GPIB Operation
8-98 PN: 10585-00001 Rev. P ML2437A/38A OM/PM
B (dB (rel)
C (dB (ref)
D (dB)
EMUL (Select emulation mod)
Syntax: B
Remarks: Set to dB units in relative mode using the present relative reference
value.
Syntax: C
Remarks: Set to dB units in relative mode using the present relative reference
value, and enable the application of the calfactor.
Syntax: D
Remarks: Set units to dBm.
Syntax: EMUL <mode>
mode:
ML24XX (Anritsu ML243xA native mode)
HP436A (Hewlett-Packard)
HP437B (Hewlett-Packard)
HP438A (Hewlett-Packard)
ML4803 (Anritsu ML4803A Series)
Remarks: Sets the GPIB emulation to emulate other types of power meters. This
command is available in any emulation mode, and resets the whole
GPIB interface when the emulation mode is changed.
When selecting GPIB emulation modes, the instrument configures itself
to the preset conditions of the instrument to be emulated. For example,
when selecting HP438A emulation, the front panel menus pass through
the presets for the HP437B (which presets sensor A to dBm) then selects
HP438A emulation (which presets sensor A to Watts).
Note: This command requires a white space between the command header
(EMUL) and the parameter <mode>. This is an ML243xA-specific
command that does not conform to the HP Emulation command format
defined at the beginning of this section.