EasyManua.ls Logo

Anritsu ML2437A - OBDSP (BNC Analog Output Display Stop Value); OBDST (BNC Analog out Display Start Value)

Anritsu ML2437A
314 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ML24xxA Native Commands GPIB Operation
8-58 PN: 10585-00001 Rev. P ML2437A/38A OM/PM
OBDSP (BNC analog output display stop value)
OBDST (BNC analog out display start value)
Syntax: OBDSP <port>,<units>,<val>
port: 1 or 2
units: W (Watts)
DB (dB)
DBM (dB)
DBUV (dBmV)
DBMV (dBmV)
val: 0 to 1 kW
-70 to 47dB
-23 to 94 dBmV
37 to 154 dBmV
Remarks: Sets up the stop value for the analog out scale of the display.
Query: OBDSP? <port>
Returned String: OBDSP <port>,<units>,<val>
Syntax: OBDST <port>,<units>,<val>
port: 1 or 2
units: W (Watts)
DB (dB)
DBM (dB)
DBUV (dBmV)
DBMV (dBmV)
val: 0 to 1 kW
-70 to 47dB
-23 to 94 dBmV
37 to 154 dBmV
Remarks: Sets up the start value for the analog out scale of the display.
Query: OBDST? <port>
Returned String: OBDST <port>,<units>,<val>