GPIB Operation ML24xxA Native Commands
ML2437A/38A OM/PM PN: 10585-00001 Rev. P 8-23
CFUADD (Add an entry pair to a cal factor table)
CFUCT (Clear cal factor table)
Syntax: CFUADD<s>,<table number>,<frequency value>[units],
<cal factor>,<cal factor units>
s: A or B
table number: 1 to number of tables supported by the sensor type
frequency value: 10 kHz to 122 GHz
cal factor: 0.07 to 150% 31.55 to –1.76 dB
cal factor units: %, PCT, DB, or DBM
Remarks: Adds an entry pair to a cal factor table. This only affects the copy of the
cal factor table stored in the memory of the power meter. Cal factors
entered with this command will be available for use by the DSP, but will
NOT be saved to the sensor until a save command (CFUSAV) is
executed. If the sensor is changed or power is lost before saving, all
changes made since the last CFUSAV will be lost.
The user must ensure that the maximum number of cal factor data pairs
entered into a table is not exceeded. Sensors with a maximum frequency
of up to 40 GHz will hold 90 pairs, while sensors with a maximum
frequency of 50 GHz will hold 110 pairs.
Related
commands:
CFUSAV
Syntax: CFUCT <s>,<table number>
s: A or B
table number: 1 to number of tables supported by the sensor type
Remarks: Clears the cal factor table to one entry for 50 MHz at 100%, but does not
clear the identity of the table. The cleared table is automatically saved
to the sensor.