ML24xxA Native Commands GPIB Operation
8-64 PN: 10585-00001 Rev. P ML2437A/38A OM/PM
OGD (Output Graph Data)
OGSD (Output Valid Samples Array (Power vs. Time mode only))
OI (Output device identification)
Syntax: OGD
Remarks: Outputs the next complete set of graph data.
The format is as follows:
OGD
<number_of_elements>,<element_1>,<element_2>,<element_n>...<\n>
The first number in the string is the number of elements to follow, and is
always 200 for the ML243xA.
Syntax: OGSD
Remarks: A power verses time chart plots the readings on a scrolling chart from
left to right. If GRCP (connect points) is on and no new data has been
received for a time slot, the graph is plotted with the same value as the
previous time slot but the data for that sample position is not marked as
valid. This command will read out an array of 1's and 0's that indicate
whether the data for that time slot is valid. For example, '1' for valid and
'0' for connecting data only.
Because the time between reading the data and reading the valid
sample data may shift the valid samples out of sync with the graph read,
it is recommended that a 'HOLD ON' be issued before reading the graph
and sample data, and a 'HOLD OFF' after. This will make sure that the
sample data and the graph data agree.
If not in Power vs. Time mode, this command will set an execution error.
Syntax: OI
Remarks: Returned format:
<Company name>,<model>,<serial>,<firmware version>