GPIB Operation HP438A Emulation Commands
ML2437A/38A OM/PM PN: 10585-00001 Rev. P 8-127
LN (Set linear units)
OC (Set calibrator state)
OI (HP Identity request)
OS (Set offset value)
Syntax: LN
Remarks: Changes the display to linear units (Watts).
Syntax: OC<state>
state:
0 (OFF)
1 (ON)
Remarks: For example: OC0 (reference calibrator state set to OFF).
Syntax: OI
Remarks: The format of the returned string is:
<Company name>,<model>,<serial>,<firmware version>
Syntax: OS<val>EN
val: -200 to +200 (dBm only)
Remarks: Specifies the offset applied to the displayed value. Values can be entered
in 0.01 dB increments.
Example: To set an offset of 10.13 to the displayed value:
OS10.13EN