Calibration and Zero Error Codes
B-12 PN: 10585-00001 Rev. P ML2437A/38A OM/PM
B-4 Calibration and Zero Error Codes
Table B-2. Error Messages
Error Message Meaning
DDDx DC_Null failure
DDxD, x=odd number DC-range 2 gain failure
DDxD, x=4-7,C-F AC_Null failure
DDxD, x=8-F Reference voltage range calibration failure
DxDD, x=odd number Range 4 failed to saturate during calibration
DxDD, x=non zero Residual Zero too high (+ve) on ranges 3-5
XDDD, x=non zero Residual Zero too high (-ve) 0n ranges 3-5