HP437B Emulation Commands GPIB Operation
8-110 PN: 10585-00001 Rev. P ML2437A/38A OM/PM
EN (Enter command)
ET (Enter data for a cal factor table)
EX (Exit cal factor table mode)
FA (Auto average)
Syntax: EN
Syntax: ET<table_number><freq_value><cal factor><terminator>
table number:
0 to 9 (F for factory table allowed when using to read a table)
freq value:
cal factor entry frequency value
cal factor:
cal factor value in percentage
terminator:
EN to terminate and entry
EX to terminate table entries
Remarks: Since the ML243xA stores the cal factor table information in the sensor,
this data must be saved to the sensor by using the ‘EX’ command or an
additional command ‘SV’, or the data could be lost. The saving of the cal
factor table data to the sensor can be done at the end of all updates to a
particular table.
Syntax: EX
Remarks: Used on the ML243xA to force a save of the cal factor table to the sensor
if the data has changed.
Syntax: FA
Remarks: Automatic Filter on. Allows the system to automatically select the filter
used to reduce the jitter in the display.
Related
commands:
FM, FH