GPIB Operation ML24xxA Native Commands
ML2437A/38A OM/PM PN: 10585-00001 Rev. P 8-31
CVSTV (V/GHz calibration factor start voltage)
DBLGHT (Battery LCD Back light mode)
DBLTIM (Auto Backlight OFF timer setting)
Syntax: CVSTV <s>,<val>[units]
s: A or B
val: -0.5 to 20.5
Remarks: Sets the start voltage of the V/GHz calibration factor settings.
Related
commands:
CVSPV, CVSPF, CVSTF
Query: CVSTV? <s>
Returned String: CVSTV <s>,<val>
Syntax: DBLGHT <mode>
mode: ON, OFF, TIMED
Remarks: Sets the mode of the LCD backlight when under Battery power.
ON = back light is ON all the time
OFF = back light is OFF all the time
TIMED = back light is on for a limited time period set by the DBLTIM
command.
Note: Although GPIB is not available under battery operation, the state of this
battery-specific parameter can be changed through this GPIB command.
Related
commands:
DBLTIM
Query: DBLGHT?
Returned String: DBLGHT <mode>
Syntax: DBLTIM <val>
val: 1.0 to 100.0 minutes
Remarks: Sets the time limit when the backlight will turn off if the DBLGHT
setting is set to TIMED.
Note: Although GPIB is not available under battery operation, the state of this
battery-specific parameter can be changed through this GPIB command.
Related
commands:
DBLTIM?
Returned String: DBLTIM <val>