GPIB Operation HP437B Emulation Commands
ML2437A/38A OM/PM PN: 10585-00001 Rev. P 8-103
*CLS (Clear GPIB status byte)
*ESE (Set the Event Status register enable mask)
Syntax: *CLS
Remarks: This command performs a status structure clear command. The event
status register and the status register are cleared except for the MAV
bit.
Syntax: *ESE<val>
val: 8-bit mask
Remarks: Event registers for the HP437B.
Bit 7: Power ON
Bit 6: N/A
Bit 5: Command error
Bit 4: Execution error
Bit 3: Device Dependent error
Bit 2: N/A
Bit 1: N/A
Bit 0: N/A
See the HP437B manual for details about the HP status registers.