ML24xxA Native Commands GPIB Operation
8-24 PN: 10585-00001 Rev. P ML2437A/38A OM/PM
CFUID (Cal factor table identity update)
CFUNITS (Cal factor display units)
CFULD (Cal factor table binary load)
Syntax: CFUID <s>,<table number>,<identity>
s: A or B
table number: 1 to number of tables supported by the sensor type
identity: Seven characters or until a message terminator will be read as
the identity.
Remarks: Updates the seven character identity string. This only affects the copy of
the cal factor table stored in the memory of the power meter. To take
affect and not be lost, the table must be saved to the sensor using the
CFUSAV command.
Query: CFUID? <s>,<table number>
Returned String: CFUID <s>,<table number>,<identity>
Syntax: CFUNITS <s>,<units>
s: A or B
units: % or PCT dB or dBm
Remarks: This command changes the display units of the calfactors between
either dB or percentage. Note that this will also set the form the data is
output over the GPIB (or RS232) when requested.
Query: CFUNITS? <s>
Returned String: CFUNITS <s>,<units>
Syntax: CFULD<s>,<table number>,<length>,<binary data>
s: A or B
table number: 1 to number of tables supported by the sensor type
length: Length of message in bytes
binary data: Same data as that received by CFURD
Remarks: Loads binary data into the cal factor table. This command will
automatically save the data to the sensor.