ON (Output number of channel readings)
OPMD (ML243xA operation mode)
Syntax: ON <c>,<val>
c: 1 or 2
val: 1 to 1000
Remarks: Readout and Power vs. Time modes only. This command returns the
specified number of readings for the specified channel. The readings are
first assembled, and then passed to the GPIB as a whole, with a line feed
character (hex 0x0a) marking the end of the string.
Example: ON 1, 9
This example will return:
–10.234, –10.234, –10.235, –10.238, -10.250, –10.270, –10.500, –10.934,
–12.234<0x0a>
Syntax: OPMD <mode>
mode: DIGIT
PROFILE
PWRTIM
SRCSWP
Remarks: This command selects the ML243xA operation mode (data collection
method).
DIGIT = digital read out of channel data
PROFILE = profile of graphic display
PWRTIM = graph of channel power versus time
SRCSWP = source sweep graphic display
To use Graph output commands, the ML243xA must be in Profile or
Power vs. Time mode. To use the output channel data commands, the
ML243xA must be in the digital readout mode.
Related
commands:
GRMD
Query: OPMD?
Returned String: OPMD <mode>