EasyManua.ls Logo

Anritsu ML2437A - SYSLNM (Saved Set Naming); SYSRD (Output the Saved Setup over the GPIB)

Anritsu ML2437A
314 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ML24xxA Native Commands GPIB Operation
8-76 PN: 10585-00001 Rev. P ML2437A/38A OM/PM
SYSLNM (Saved set naming)
SYSRD (Output the saved setup over the GPIB)
Syntax: SYSLNM <store number>,<text>
store number: 1 to 10
text: text string
Remarks: This command allows the saved setups to have text associated with
them rather than just the ‘USED’ and ‘NOT USED’ text.
Query: SYSLNM? <store number>
Returned String:
SYSLNM <store number>,<store name>
If a store number of 0 is used, then all the store titles will be output in
the form:
SYSLNM 1,<store 1 name>,2,<store 2 name>, … ,10,<store 10 name>
Syntax: SYSRD <store number>
store number: 0 (current setup) or 1 to 10 saved stores
Remarks: Requests that the saved stored setup is output over the GPIB. This is a
BINARY output that allows the stored setup to be programmed into
other ML243xA power meters and stores via the SYSLD command. If a
request for a store number that has not had a setup stored into it is
made, an execution error event will be set in the Event Status Register
(ESR).
The output is in the form:
SYSRD<ws><#><num_digits><number num_digits long>, <binary
data>
<num_digits> = Number of following digits giving the number of bytes of
binary data.
<number num_digits long> = A number num_digits long giving the
number of bytes of binary data.
<binary data> = Saved setup.