GPIB Operation ML24xxA Native Commands
ML2437A/38A OM/PM PN: 10585-00001 Rev. P 8-33
DPEAK (Peak meter display)
DTRGD (Display Trigger Delay)
Syntax: DPEAK <mode>
mode: A, B, A&B, OFF
Remarks: Turns the peak meter display on or off for each channel.
A = Sensor A only
B = Sensor B only
A&B = Sensors A and B displayed at the same time
OFF = Turns the peak meter display off.
The peak meter display range covers 12 dB. When above the displayed
maximum or below the displayed minimum, the range is switched by 10
dB in the appropriate direction.
Note that in the event that the channel is displaying an alternative
measurement (e.g., external volts from the rear panel BNC) the peak
meter continues to represent the Sensor A and/or B data. This is very
useful for monitoring an external voltage on the meter, while peaking up
a response being monitored by a sensor (e.g., RF output).
Query: DPEAK?
Returned String: DPEAK <mode>
Syntax: DTRGD <val>
val: 0.0 to 7.0 s
Remarks: The delay time from the trigger point to when the profile starts to be
drawn.
Example: DTRGD 1.25MS Sets the display trigger delay to 1.25 ms.
Query: DTRGD?
Returned String: DTRGD <val>