ML24xxA Native Commands GPIB Operation
8-52 PN: 10585-00001 Rev. P ML2437A/38A OM/PM
LLIM (Set Low limits)
LLIMS (Turn on/off low limits)
Syntax: LLIM <c>,<val>
c: 1 or 2
val:
Units Min Max
dBm -270 220
dBmV -53.00 147.00
dBV 7.00 207.00
Watts 0.0 50.0
Volts 0.0 20.0
Remarks: Sets the low limit. The LLIMS command turns the limits on and off. It is
not necessary to enter the units as the limit value is checked against the
displayed value.
Therefore, if the limits have been set for –10 dBm (LLIM 1, –10) and the
display units are subsequently changed from dBm to Watts, the system
still checks for the reading to rise above –10, even though the display
units type has been changed.
Related
commands:
LLIMS
Query: LLIM? <c>
Returned String: LLIM <c>,<val>
Syntax: LLIMS <c>,<state>
c: 1 or 2
state: ON or OFF
Remarks: The LLIMS command turns the limits on and off.
Related
commands:
LLIM
Query: LLIMS? <c>
Returned String: LLIMS <c>,<state>