HP438A Emulation Commands GPIB Operation
8-126 PN: 10585-00001 Rev. P ML2437A/38A OM/PM
KB (Calibration factor)
LG (Set log units)
LH (Set high limit)
LL (Set low limit)
LM (limits check state)
Syntax: KB<val><terminator>
val: 1.0 to 150.0%
terminator:
%
PCT
EN
Remarks: The calibration factor compensates for mismatch losses and effective
efficiency over the frequency range of the power sensor.
Example: KB99.9%
KB99.9PCT
KB99.9EN
Syntax: LG
Remarks: Changes the display to log units (dBm).
Syntax: LH<val>EN
val: -200 to +200 (dBm only)
Remarks: Sets the high limit.
Example: LH30.00EN
Syntax: LL<val>EN
val: -200 to +200 (dBm only)
Remarks: Sets the low limit.
Example: LL20.00EN
Syntax: LM<state>
state: 0 (off) or 1 (on)
Remarks: Turns limit checking on or off.