ML24xxA Native Commands GPIB Operation
8-28 PN: 10585-00001 Rev. P ML2437A/38A OM/PM
CHRES (Set channel decimal pint resolution)
CHUNIT (Set Channel units)
CONT (Continue)
Syntax: CHRES <c>,<val>
c: 1 or 2
val: 1 to 3
Remarks: Set the number of decimal places displayed for the specified channel.
For example, specifying CHRES 1, 1 would yield a display of 1.5 dBm;
CHRES 1, 2 would yield 1.47 dBm; CHRES 1, 3 would yield 1.468 dBm.
If the number to be displayed is too large for the number of decimal
places selected, the decimal places displayed will be reduced so that the
display value can be shown.
Query: CHRES? <c>
Returned String: CHRES <c>,<val>
Syntax: CHUNIT <c>, <units>
c: 1 or 2
units: W (Watts), DBM (dB), DBUV (dBmV), DBMV (dBmV)
Remarks: DBM 0dB is equal to 1mW readout mode
W = Watts readout mode
V = Volts readout mode. This selection is automatically made when the
channel input configuration is set to External volts (EXT V).
DBUV = dBmV, 0dB is equal to 1mV in readout mode.
Query: CHUNIT? <c>
Returned String: CHUNIT <c>,<units>
When the channel input configuration is set to External volts (EXT V),
the returned units are always volts, irrespective of what units have been
set.
Syntax: CONT
Remarks: This command will allow the system to continue the startup sequence if
there are self test failures other than DSP errors.
Related
commands:
STERR, START