ML24xxA Native Commands GPIB Operation
8-74 PN: 10585-00001 Rev. P ML2437A/38A OM/PM
: SSSS = Sensor B average number (0000 if ML2437A). For Profile and
Source Sweep modes, this number is between 1 and 512. For digital
Readout or Power vs. Time modes, the values are either 1 to 512 or, if in
AUTO averaging mode, 513 to 1024.
T = Sensor A low level average: '0' = OFF, '1' = Low, '2' = Medium, '3' =
High
U = Sensor B low level average: '0' = OFF, '1' = Low, '2' = Medium, '3' =
High.
V = Sensor A zeroed status: '0' = Not zeroed, '1' = Zeroed.
W = Sensor B Zeroed status: '0' = Not zeroed, '1' = Zeroed.
X = GPIB trigger mode: '0' = TR0 hold ON, '1' = Free run.
Y = GPIB group trigger mode: '0' = GTO, '1' = GT1, '2' = GT2.
Z = Calibrator state: '0' = OFF, '1' = ON.
1 = GPIB DISP command status: '0' = OFF, '1' = ON.
2 = GPIB FAST status: '0' = OFF, '1' = ON.