Power-On Procedure Front Panel Operation
5-4 PN: 10585-00001 Rev. P ML2437A/38A OM/PM
Failure and warning messages that can be displayed on the front panel are:
• Flash code csum
• Personality csumVolatile RAM
• Cal data csum
• Non-Vol RAM
• Software version - this is only a warning
• Current Setup
• Saved Setups
• Secure - Mem clear - this is only a warning
• Display
• Keyboard
• DSP error # - followed by a 4-digit hexadecimal error code
If an error, other than a DSP error, is encountered, the text: “Press ANY key to continue” is
displayed at the bottom of the screen. If only warnings are encountered, nothing is displayed
at the bottom of the screen, and the unit continues to initialize.
If a DSP error occurs, the text, "Restart unit. If error persists, contact Service Center." is
displayed and the unit will halt the self test. Make a note of the hexadecimal error code
displayed and contact your nearest Anritsu Service Center.
Self test error messages are also available over the GPIB, as a self test status string (see
STERR command in chapter 7).