Overview
Applied Biosystems 7900HT Fast Real-Time PCR System and SDS Enterprise Database User Guide 6-19
Automatic Outlier Removal
About Outlying
Replicate Wells
For any PCR, experimental error may cause some wells to amplify insufficiently or
not at all. These wells typically produce C
T
values that differ significantly from the
average for the associated replicate wells. If included in the calculations, the C
T
values from these wells indicate an incorrect level of relative gene expression by
skewing the average for the replicate group.
About the
Automatic Outlier
Removal
The SDS software offers algorithmic identification and removal of outlying data for
studies consisting of replicate populations of three or more wells. The statistical
method used by the software is based on Grubbs outlier removal (also known as the
Maximum Normalized Residual Test), which permits the exclusion of a single outlier
in a population consisting of as few as three replicates.
The software applies the Grubbs tests at different stages in the ∆∆C
T
calculation,
depending on the type of endogenous control used in the study. For non-multiplex
studies, the software removes outliers from replicate groups immediately after
calculating threshold cycles (C
T
). For multiplex studies, the software applies the
algorithm following the ∆C
T
calculation. Also, if an apparent outlier is within 0.25
C
T
s of the mean for the associated replicate group, the software does not remove it.
Note:
The outlier removal feature is optional and can be activated from the Analysis
Settings dialog box (see
page 6-29
). Outliers can also be identified and eliminated
manually as explained on
page 6-49
.
In addition to the Grubbs tests, the algorithm used by the
SDS software
features
additional rules for dealing with SDS data. Unless the majority of samples in a
replicate population are at maximum cycle, the algorithm ignores max cycle wells
(wells with C
T
s equal to or greater than the Maximum cycle value).
Regarding
Amplification
Beyond
Reproducible
Limits
The experimental reproducibility of C
T
values for relative quantification assays
decreases as the starting copy number approaches extremely low levels (less than
100 copies). Therefore, PCR targets with low starting copy number and subsequent
high C
T
values may have significantly diminished statistical reproducibility. To avoid
problems with reproducibility for low-expression targets, the limits of reproducibility
must be determined experimentally for each relative quantification assay (all targets
and the endogenous control).

Table of Contents

Other manuals for Applied Biosystems 7900HT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7900HT and is the answer not in the manual?

Applied Biosystems 7900HT Specifications

General IconGeneral
BrandApplied Biosystems
Model7900HT
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals