Analysis Checklist
Applied Biosystems 7900HT Fast Real-Time PCR System and SDS Enterprise Database User Guide 6-9
Analysis Checklist
Workflow
Overview
1. Open the plate document (see page 6-10).
2. Configure the analysis settings for the run (see page 6-11).
3. Analyze the run data (see page 6-12).
4. For each detector that you configured to determine the baseline and threshold
values (see page 6-46):
Automatically Verify that the baseline and threshold values for the detector
were set correctly by the software.
Manually Set the baseline and threshold values for the detector
5. Visualize outliers and eliminate any outlying amplification from the run data
(see page 6-49).
6. View the results of the absolute quantification run (see page 6-14).
7. Choose from the following post-analysis options (see page 6-52):
Reanalyze the run data.
Adjust the display settings for the results table, plate grid, and plate
document plots.
Print elements of the plate document.
Export the plate document results table or plots.
Save the analysis.