Eliminating Outlying Amplification
Applied Biosystems 7900HT Fast Real-Time PCR System and SDS Enterprise Database User Guide 6-49
Eliminating Outlying Amplification
Overview
For any PCR, experimental error may cause some wells to amplify insufficiently or not
at all. These wells typically produce C
T
values that differ significantly from the average
for the associated replicate wells. If included in the absolute or relative quantification
calculations, these outliers can potentially result in erroneous measurements.
Visualizing
Outliers
To ensure precise relative quantification, replicate groups must be carefully
examined for outlying wells. The C
T
vs. Well Position view of the Amplification plot
allows you to examine each set of replicate wells for outliers.
1. Select the Results tab.
2. If analyzing a Relative Quantification study, select a plate in the Plates field and
deselect all wells in the plate grid. Repeat for each plate in the Plates field.
When you are finished, the Amplification plot should not display any data.
3. If eliminating outliers from a:
Absolute Quantification Plate Document
Select all wells of the plate by clicking the upper-left corner of the plate grid.
Relative Quantification Multiple Plate Document
a. Select a plate in the
Plates
field to view.
b. Select all wells of the plate by clicking the upper-left corner of the plate grid.
4. In the Plot drop-down list, select Ct vs. Well Position.
5. In the Well versus Threshold Cycle Amplification plot, verify the uniformity of
each replicate population by comparing the groupings of C
T
values for the wells
that make up the set.
6. In the Detector drop-down list, select another detector, and repeat steps 3
through 5. Repeat until each detector has been screened for outliers.
7. If analyzing a Relative Quantification Multiple Plate Document, deselect the
wells of the current plate, then repeat steps 3 through 6 for every plate in the
study.
8. Are outliers present on the plate?
Yes If outliers are present for any detector, eliminate the outlying wells from
the analysis by omitting them as explained in “Eliminating Wells from the
Analysis” on page 6-50.
No If no outliers are present, do one of the following:
If analyzing an Absolute Quantification Plate Document, view the results of
the run displayed within the Standard Curve plot and Results table.
Are outliers present? Then…
Yes Record the well numbers of the outlying wells, then go to
the next step.
No Go to the next step.

Table of Contents

Other manuals for Applied Biosystems 7900HT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7900HT and is the answer not in the manual?

Applied Biosystems 7900HT Specifications

General IconGeneral
BrandApplied Biosystems
Model7900HT
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals